Important: phase correction is no longer allowed after baseline correction.
Right click to set/unset pivot point for 1st order phase correction.
Hold shift (1 degree per step) or control (0.1 per step) key, then use mouse wheel to adjust phase.
Important: phase correction is no longer allowed after baseline correction.
Right click to set/unset pivot point for 1st order phase correction.
Hold shift (1 degree per step) or control (0.1 per step) key, then use mouse wheel to adjust phase.
A appreciate baseline correction can only be defined by the user, based on the experimental conditions. That is, the user need to define the smoothness parameter and positive spectrum parameter then apply the baseline correction after satisfying result.
The smoothness parameter is the weight of the smoothness constraint in the baseline estimation. The larger the value, the smoother the baseline.
The positive spectrum parameter is the weight of the constraint that the baseline should be below the spectrum. The larger the value, the less the baseline will be allowed above the spectrum (to accommodate noise).